Knowledge

Installation of thermocouples

Jul 02, 2024 Leave a message

1) Thermocouples should not be installed next to the furnace door or near the heating object, nor should they be installed in a location with a strong magnetic field.

2) The thermocouple should be installed vertically as much as possible to prevent the deformation of the protective tube at high temperatures; When the fluid needs to be mounted at an angle, the thermocouple should be supported with an appropriate bracket.

3) When installing the thermocouple with porcelain protective tube, the position that does not hinder the movement of the workpiece should be selected, and the installation hole should have enough thermal denaturation space to avoid the porcelain tube breaking; At the same time, it should be avoided to avoid rapid cold and heat to prevent the porcelain tube from bursting.

4) The thermocouple terminal should not touch the surface of the object to be measured, and the temperature of the cold terminal should generally not exceed 70°C, and it should be kept as stable as possible.

5) The outlet hole of the junction box should be in a horizontal position until it is downward to prevent water vapor, dust, and stolen goods from falling into the junction box.

6) The insertion depth of the thermocouple is determined according to the actual needs, but the length immersed in the measured medium should be greater than 8-10 times the diameter of its protective tube.

7) For thermocouples that measure the temperature of solid surfaces or molds, the measuring end part must be in good and close contact with the measured object, and the measured surface should be kept clean and clean.

8) When the thermocouple is wired, the junction box cover should be opened first, and then the compensation wire should be penetrated through the rubber plug from the outlet hole of the junction box, and the polarity should not be reversed.

9) In order to protect the compensation wire from external mechanical damage, and to avoid the external magnetic field affecting the electronic display instrument, the compensation wire should be shielded or loaded into the grounded steel pipe if necessary, and the compensation wire should not have twists and turns.

10) A thermocouple cannot be connected to two display meters at the same time, except after installing a temperature transmitter.

11) When the fixing ferrule device is used for the first time, the insertion depth is adjustable, and it should be tightened immediately after adjustment, but attention must be paid to the appropriate force to ensure that the pressure seal is met without damage. Once tightened, the insertion depth is no longer adjustable.

12) Movable ferrule device, the insertion depth is also adjustable, but it cannot withstand pressure. If appropriate auxiliary sealing packing (such as asbestos rope, etc.) is added, it can also withstand a certain sealing pressure.

When the longer armored thermocouple is arranged and installed along a non-straight path, it should be noted that the bending radius should not be less than 5 times the diameter of the thermocouple, and there should be no mechanical damage, no right-angle sharp bend, and the clamp should be fixed if necessary.

Send Inquiry